Rotary flow regulators have a standardized flow rate stability of +/- 10% over time (ISO8536-14 standard), and are therefore considered lt;precision lt; regulators. It should be noted that many factors influence flow regulation (viscosity, height, blood pressure, etc...). Droplet counting and regular monitoring are therefore still good practice when using these precision regulators. This is why this rotary flow regulator is offered without graduation, to enable good compliance with these good practices. - After purging the line, and before connecting the device to the patient's vascular access, turn the flow regulator to the lt;off lt; position. Contact us to obtain the good use sheet.
Extension line with :
- High-quality tubing: flexible, anti-plication PVC
- Precision rotary flow regulator: Smooth, easy adjustment - Constant, reliable flow
- Optimum technical features:
Fixed and mobile Luer Lock tips
Pressure resistance: 50 KPA (0.5 Bars - 7.25psi)
Non-graduated flow regulator: No need to count drops
Made in China.
To ensure optimum adjustment, please follow the purging protocol below :
- Purge the line with the flow regulator in the lt;open lt; position (delivered in the lt;open lt; position).
Contact with Chlorhexidine 0.5% or 2% in Isopropyl alcohol
Isopropyl alcohol at 70°, with no evaporation time, and while the luer lock
under physical stress, presents a significant risk of luer lock
of luer lock cracking, and should be avoided wherever possible.
In line with SF2H (2019) recommendations, disinfecting connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective at disinfecting inert materials and much less likely to damage plastics.
Indications
This device enables precise gravity administration of injectable parenteral preparations: drugs, solutions, parenteral nutrition.
Do not use with high-risk drugs, drugs with narrow therapeutic margins (catecholamines, morphine, heparin, insulin, potassium chloride, anticancer drugs requiring infusion pump administration, and all pharmaceutical specialties whose R.C.P. includes the requirement for administration by infusion pump or syringe pump, or via an infusion set fitted with a 0.2µm filter).
Source : OMEDIT CENTRE -- Les bonnes pratiques de perfusion : Module lt; La précision du débit de perfusion lt;.
Dispose of in accordance with local regulations.
Used materials
Cap: PE
Female Luer Lock: PC
Tube: PVC
Flow regulator: ABS / TPE
Luer Lock male: ABS
Cap: PE
Labeling
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
medical device
sterile
conformity
Transport conditions
Protected from light and humidity
Conditioning | Parts quantities | Dimensions in cm | Weight in kg | Ean 13 of conditioning |
UNIT | 1 | 21 x 13 x 0 | 06972620110527 | |
BAG | 25 | |||
CASE | 150 | 39 x 42 x 31 | 16972620110524 |